RT Machine Verification Service Class

The RT Machine Verification Service Class defines a service that uses the DIMSE N-CREATE, N-SET, N-DELETE, N-ACTION, N-EVENT-REPORT and N-GET services to facilitate the independent verification of geometric and dosimetric settings on a radiation delivery system prior to delivery of treatment.

Supported SOP Classes

UID

SOP Class

1.2.840.10008.5.1.4.34.8

RTConventionalMachineVerification

1.2.840.10008.5.1.4.34.9

RTIonMachineVerification

DIMSE Services

DIMSE Service

Usage SCU/SCP

RT Conventional Machine Verification

RT Ion Machine Verification

N-CREATE

Mandatory/Mandatory

N-SET

Mandatory/Mandatory

N-GET

Mandatory/Mandatory

N-ACTION

Mandatory/Mandatory

N-DELETE

Mandatory/Mandatory

N-EVENT-REPORT

Mandatory/Mandatory

Statuses

N-ACTION Statuses

Code (hex)

Category

Description

0x0000

Success

Success

0x0112

Failure

No such SOP Instance

0x0114

Failure

No such argument

0x0115

Failure

Invalid argument value

0x0117

Failure

Invalid object instance

0x0118

Failure

No such SOP Class

0x0119

Failure

Class-Instance conflict

0x0123

Failure

No such action

0x0124

Failure

Refused: not authorised

0x0210

Failure

Duplicate invocation

0x0211

Failure

Unrecognised operation

0x0212

Failure

Mistyped argument

0x0213

Failure

Resource limitation

RT Machine Verification N-ACTION Service Statuses

Code (hex)

Category

Description

0xC112

Failure

Applicable Machine Verification instance not found

N-CREATE Statuses

Code (hex)

Category

Description

0x0000

Success

Success

0x0105

Success

No such attribute

0x0106

Success

Invalid attribute value

0x0107

Success

Attribute list error

0x0110

Success

Processing failure

0x0111

Success

Duplicate SOP Instance

0x0116

Success

Attribute value out of range

0x0117

Success

Invalid object instance

0x0118

Success

No such SOP Class

0x0120

Success

Missing attribute

0x0121

Success

Missing attribute value

0x0124

Success

Refused: not authorised

0x0210

Success

Duplicate invocation

0x0211

Success

Unrecognised operation

0x0212

Success

Mistyped argument

0x0213

Success

Resource limitation

RT Machine Verification N-CREATE Service Statuses

Code (hex)

Category

Description

0xC221

Failure

The Referenced Fraction Group Number does not exist in the referenced plan

0xC222

Failure

No beams exist within the referenced fraction group

0xC223

Failure

SCU already verifying and cannot currently process this request

0xC227

Failure

No such object instance - Referenced RT plan not found

N-DELETE Statuses

Code (hex)

Category

Description

0x0000

Success

Success

0x0110

Failure

Processing failure

0x0112

Failure

No such SOP Instance

0x0117

Failure

Invalid object Instance

0x0118

Failure

Not such SOP Class

0x0119

Failure

Class-Instance conflict

0x0124

Failure

Not authorised

0x0210

Failure

Duplicate invocation

0x0211

Failure

Unrecognised operation

0x0212

Failure

Mistyped argument

0x0213

Failure

Resource limitation

N-EVENT-REPORT Statuses

Code (hex)

Category

Description

0x0000

Success

Success

0x0110

Failure

Processing failure

0x0112

Failure

No such SOP Instance

0x0113

Failure

No such event type

0x0114

Failure

No such argument

0x0115

Failure

Invalid argument value

0x0117

Failure

Invalid object Instance

0x0118

Failure

No such SOP Class

0x0119

Failure

Class-Instance conflict

0x0210

Failure

Duplicate invocation

0x0211

Failure

Unrecognised operation

0x0212

Failure

Mistyped argument

0x0213

Failure

Resource limitation

N-GET Statuses

Code (hex)

Category

Description

0x0000

Success

Success

0x0107

Warning

SOP Class not supported

0x0110

Failure

Processing failure

0x0112

Failure

No such SOP Instance

0x0117

Failure

Invalid object instance

0x0118

Failure

No such SOP Class

0x0119

Failure

Class-Instance conflict

0x0122

Failure

SOP class not supported

0x0124

Failure

Not authorised

0x0210

Failure

Duplicate invocation

0x0211

Failure

Unrecognised operation

0x0212

Failure

Mistyped argument

0x0213

Failure

Resource limitation

RT Machine Verification N-GET Service Statuses

Code (hex)

Category

Description

0xC112

Failure

Applicable Machine Verification instance not found

N-SET Statuses

Code (hex)

Category

Description

0x0000

Success

Success

0x0105

Failure

No such attribute

0x0106

Failure

Invalid attribute value

0x0110

Failure

Processing failure

0x0112

Failure

SOP Instance not recognised

0x0116

Failure

Attribute value out of range

0x0117

Failure

Invalid object instance

0x0118

Failure

No such SOP Class

0x0119

Failure

Class-Instance conflict

0x0121

Failure

Missing attribute value

0x0124

Failure

Refused: not authorised

0x0210

Failure

Duplicate invocation

0x0211

Failure

Unrecognised operation

0x0212

Failure

Mistyped argument

0x0213

Failure

Resource limitation

RT Machine Verification N-SET Service Statuses

Code (hex)

Category

Description

0xC224

Failure

Referenced Beam Number not found within the referenced Fraction Group

0xC225

Failure

Referenced device or accessory not supported

0xC226

Failure

Referenced device or accessory not found within the referenced beam